SECTION 16 b.

TROUBLE SHOOTING CONT.

WILL NOT READ VOLTAGES ON ITS “0-33Vdc In”

INPUT JACK

Consult the suggested wiring diagrams in section 14j referring to

the proper voltage measurement setup

With the unit powered up ensure that the input voltage is selected

from the front keypad (refer to manual) by pressing the “mA/V”

key in the lower left hand corner of the key pad. Multiple presses

of the key will rotate through” Volts”, “mA”, and pressure only

(blank displayed in the middle of the LCD display). If voltage still

isn’t displayed when selected:

Remove the HHC from the external wiring system and connect

a DC voltage source (0-33 Volts DC Max) between the “0-33Vdc

In” jack (+) and the “COM” jack (-) located at the top of the unit.

Select “volts” using the front keypad “ma/V” key and monitor the

display. The input voltage should be displayed.

If the voltage source used is faulty you can use the 24 volt loop

supply of the HHC to check the input voltage jack. Ensure that the

loop supply is operational (only valid for non- I.S. units) _ connect

the “24Vdc Out” jack to the “0-33Vdc In” jack and read the voltage

on the display _ Note that the COM connection is not required

here as they share this connection internally to the PTE-2.

Note: The power supply of a non-intrinsically safe base unit exhibits

a leakage voltage of approximately 1 volt in the off condition. This

is normal.

If the voltage still cannot be read, consult the factory

WILL NOT READ CURRENT INPUT INTO THE

“0-22Ma In” JACK

Consult the manual’s circuit diagram section 14j, referring to the

proper current measurement setup

With the unit powered up ensure that the input current measure-

ment is selected from the keypad by pressing the “mA/V” key in

the lower left hand corner of the key pad. Multiple presses of

the key will rotate through” Volts”, “mA”, and pressure only (blank

displayed in the middle of the LCD display). If current still isn’t

displayed when selected:

Remove the HHC from the external system and connect an iso-

lated direct current source (0-22 mA Max) between the “0-22mA

In” jack (+) and the “COM” jack(-) located at the top of the unit.

Select “mA” using the keypad “mA/V” key and monitor the display.

The input current should be displayed, if the current is not dis-

played:

Using a precision voltage source input a 200mv signal (+) pos-

itive lead into the “0-22mA In” jack and the negative side of the

voltage source (-) to the “COM” jack. This will simulate a 20ma

signal. If the current still is not displayed consult the factory

WILL NOT DISPLAY PRESSURE MODULE DATA

Remove the module from the base unit and then re-install the

module making sure it is fully seated in its base unit cavity. The

modules are hot swappable and upon reinsertion the pressure

should be displayed after the units re-booting process. If no pres-

sure value is displayed:

Turn off power to the HHC by pressing the green power key.

Remove the pressure module in question. Look into the vacated

module cavity in the base unit and inspect the interface con-

nector pins inside the cavity. Ensure that none of the connector

pins are bent or damaged. In the unlikely event one or more of

the pins are bent, carefully straighten the pin(s) then re- insert

the module making sure that the locking tab is fully seated in its

mating hole at the bottom of the base unit. Turn the base unit

back on, pressure should be displayed after the units booting pro-

cess. If pressure is not displayed, try another pressure module if