Press left and RIGHT ARROWs to position cursor beneath

desired numeric character
Press UP and DOWN ARROWs to scroll numeric values.
When desired numeric value is displayed in proper decimal

position
Press left or RIGHT ARROW to move cursor beneath the next

digit to be programmed
When desired Full scale value is displayed on screen for example
00006.00000
Press ENTER to accept
HHC will go back to Percent Error MENU and advance to the next

programmable parameter.
Use DOWN or UP ARROW to highlight “Enter % Error Mode”
Press ENTER
The HHC is now actively measuring the Percent Error of the

device under test.

The HHC is equipped with firmware that can add or subtract the

measured pressure values from two installed Quick-Select pres-

sure modules. This allows for measurement of what are com-

monly referred to as “high-line” or elevated static differential pres-

sures. The ability to add the measured values of the two installed

Quick-Select pressure modules allows for the addition of a mea-

sured barometric pressure from an absolute pressure module to

the measured pressure from a gauge or differential pressure. This

capability can use gauge measurement Quick-Select pressure

modules for absolute pressure measurement.
Accuracy of Dual Module Measurements
The accuracy of measurements made using the difference or

sum of two modules is a function of the full scale range of the

Quick-Select pressure modules used and, in the case of differ-

ential pressure measurements, the magnitude of the differential

pressure range.
For dual sensor summation measurements:
The worst case inaccuracy would be the sum of the inaccuracies

of the two Quick-Select modules used for the measurement. For

example, if the measurement is made with two +0.1% full scale

modules the maximum inaccuracy would be +0.2%. In actual

use, it is generally acceptable to use the Root Sum of the Square

(RSS) method of determining the accuracy of a multiple compo-

nent measurement system. Using the RSS method the accuracy

of the dual module summation is approximately +0.14% of span.

SECTION 15 h.

PERCENT (%) ERROR FUNCTION

CONT.

SECTION 15 i.

DUAL MODE FUNCTION

Output FS

00006.00000

^

Enter to accept

Esc to cancel

50.000

In(PSi)

Out(volts)

% error

2.423

0.38