Set log interval – Ashcroft ATE-2 - Handheld Calibrator User Manual

Page 34

Advertising
background image

32

-0.11

inH20

0.000

volts

-0.011

PSi

Auto–logging

SECTION 15 f. iii.

Initiate Auto Datalog

SECTION 15 f. iv.

Initiate Manual Datalog

Set log interval

Desired: 00001.0 sec

Actual: 00001.2 sec

Enter to accept

Esc to cancel

“^ “Beneath a digit indicates which character in the time interval

name is being configured.
Press UP or DOWN ARROW to scroll through numeric values
Use left or RIGHT ARROW to move cursor to next desired

position.
When the desired time interval in seconds is displayed, press

ENTER
The HHC will return to Measurement Mode.
*NOTE*
Note: When programming, the display will show “Desired” interval

and “Actual” interval. Actual interval may not match the desired

value. The minimum sample rate changes are based on the

number of inputs installed in the base unit. The sample rate

increase 0.1 seconds per input (0.1, 0.2 or 0.3 seconds). The

actual interval must be an even multiple of this value. For exam-

ple, if there are two pressure modules and a voltage input active,

the minimum sample rate is 0.3 seconds or even multiples of 0.3

seconds. If the desired value entered is 0.1 seconds the actual

value will remain 0.3 seconds.

Initiate Automatic Datalog

To initiate an automatic Datalog after programming desired chan-

nel tag names and logging interval press DATA STORE key. The

display will show a message saying “Auto-logging”

To disable auto logging press either DATA STORE key or ESC key.

Initiate a Manual Datalog
After programming desired channel tag names
Press MENU key

Use DOWN or UP ARROW to highlight “Datalog…”

Press ENTER

SECTION 15 f. ii.

Program Auto Datalog Time

Cont.