The target Full Scale Pressure cardinal value and Min and

Max allowable limits will be displayed.

Use a pressure standard to apply target pressure value to a

pressure module. If the pressure standard does not achieve

exactly the target cardinal pressure value, actual target pres-

sure can be edited within the Min and Max limits. This value is

the pressure reading which is displayed on the external pres-

sure calibration standard.

If the cardinal Full Scale Pressure value cannot be achieved

exactly by the pressure standard:
a. Press left and RIGHT ARROWs to position cursor beneath

desired character

b. Press UP and DOWN ARROWs to increment numeric char-

acters

c. When the pressure value displayed on HHC matches the

value displayed on the pressure standard press ENTER

The display will read “Acquiring Cal Data…Wait while system

is busy”. After Full Scale Cal is complete (several seconds)

the screen returns to the Calibration Menu with “Cal Linear-

ity” selected. At this time the operator can proceed with “Ca

Linearity”, return to “Zero Cal”, or “Cal Full Scale”, or save the

Zero Cal & Full Scale Cal by selecting “Apply new cal” and

pressing “ENTER”.

6. Select “Cal Linearity” and press “ENTER”.

When calibrating linearity, the HHC will automatically choose

the number and pressure value of data points based on the

module’s full scale pressure range and its base accuracy

specification. The number of equidistant pressure values

required for linearity calibration is listed below:

0.025% & 0.05% enhanced accuracy

20 points (approximately 10% increment)

0.025% - 0.07% standard accuracy

11 points (approximately 10% increment)

0.1% standard accuracy

6 points (approximately 20% increment)

0.25% standard accuracy

5 points (approximately 25% increment)

The first linearization point target pressure and min and max

allowable limits in default engineering units will be displayed.

Using the pressure standard, apply the target pressure to the

module. If the pressure standard does not achieve exactly

the target cardinal pressure value, actual target pressure can

be edited within the Min and Max limits. Actual value is the

pressure reading which is displayed on the external pressure

calibration standard.

If the cardinal linearization pressure value cannot be achieved

exactly by the pressure standard:

a. Press left and RIGHT ARROWs to position cursor beneath

desired character

b. Press UP and DOWN ARROWs to increment numeric

characters

7. When the pressure value displayed on HHC matches the value

displayed on the pressure standard press ENTER” The system

will respond with “Acquiring Cal Data…Wait while system is busy”.

NOTE: The first linearization point pressure is equivalent

to the first pressure point after Zero for all pressure types,

except Compound. For the Compound ranges, the first cali-

bration point will be an equivalent to the lowest negative pres-

sure of the range.

Upon completion of the first calibration point inquiry, the HHC

will respond with a next calibration point screen and continue

through the process until the last calibration point is complete.

The last calibration point will be a pressure point one step

before the Full Scale pressure. Repeat steps 6 and 7 as many

times as required based on accuracy as noted above.

SECTION 15 o.

PRESSURE MODULE CALIBRATION

CONT.