SECTION 14 e.

BACK LIGHT FUNCTION

SECTION 14 d.

ZERO FUNCTION

During any programming operation pressing the ESC key will

cause the HHC to exit any special display mode or, step back one

programming level if in Programming Mode.

PORT SELECT function is used in several programming appli-

cations including zero, min/max, tare and damping. This allows

the user to choose between installed modules to which an action

is to be performed. PORT SELECT is enabled following the ini-

tialization of a desired function such as zero. The selected mod-

ule(s) will be indicated by having a rectangular box surrounding

the measurement indication. Pressing the PORT SELECT key

repeatedly will allow the user to select the Channel 1 , Channel

2 or both channels if applicable. When desired channel (s) are

selected the ENTER key is pressed to finalize the selection.

The default Data Storage mode is manual and is described in

section 14g and 15f. The HHC is capable of storing pressure,

temperature, voltage or current data for any combination of

installed modules, voltage or current inputs that are active, up

to three simultaneous data points. Each data record will include

a time stamp and up to three measurement values. Headers in

the Datalog report label columns with date, user-programmable

module description (default values are Channel 1, Channel 2,

Ext.) and the engineering unit of measure. Up to 16,000 records

can be recorded. Data records are grouped in Datalogs which

can be manually initiated or automatically logged at user defin-

able time intervals between 100 milliseconds and 24 hours.

Utilized to remove zero offset, up to 10% of full scale range, when

zero pressure is applied.
1. Press the ZERO key. PORT SELECT MENU is entered with

the Channel 1 module selected.

2. Use the PORT SELECT key to select Channel 1, Channel 2 or

both Channels.

3. Press ENTER to zero the selected module or modules, or

press ESC to cancel and return to Measurement Mode

A module must be installed in the HHC base unit to be zeroed.

Additionally, it must be a pressure module, and must be within

10% of full scale of its factory programmed zero.
In the event an attempt is made to zero a module that is absent,

or which is not a pressure module a warning will be displayed in

the notice area stating:
“CH(X) not a pressure mod”.
In the event an attempt is made to zero a module that is more

than 10% of full scale from its factory programmed zero, a warn-

ing will be displayed in the notice area stating:
“CH(X) zero out of bounds”

Press the Back Light key to toggle the light on/off. A user pro-

grammable back light auto off timer can be used to preserve

battery life. Press any key to reset the backlight auto off timer.

SECTION 14 a.

ESC FUNCTION

SECTION 14 b.

PORT SELECT FUNCTION

SECTION 14 c.

DATA STORAGE FUNCTION

–0.12

in H 2 0

-0.012

PSi