Ashcroft ATE-2 - Handheld Calibrator User Manual

Page 39

Advertising
background image

37

The Percent Error Function allows the user to display the output

from a device under test in terms of percent error of the full scale

range of that device. This function applies to the calibration of

both temperature and pressure devices. Percent error mode

eliminates the need for manual calculations to determine the level

of accuracy of the device under test.
The HHC is capable of comparing two measurements in several

configurations
• Pressure or temperature input to current output device, such as

a transducer with 4-20 mA output

• Pressure or temperature input to voltage output device, such as

a transducer with 0-10 Vdc output

• Pressure to pressure device, such as a pneumatic controller
• Current input to pressure output device such as a pressure

regulator with 4-20 mA input signal

All input and output parameters can be programmed by the

operator.
Datalogging is possible while using the HHC in Percent Error

Mode. Datalogs will include only the actual values measured. Dat-

alogs will not include the percent error calculation as displayed

on the HHC screen.
For example, if the HHC is being used to calibrate a 0-100 psi

transmitter with 4-20 mA output the HHC will display the following

data:
• Input pressure, as measured by Quick-Select pressure module

in selected engineering units that is being applied to the device

under test. This is the standard against which the device under

test is being compared.

• Output, from the transducer under test in output units (milli-

amps)

• Percent error, calculated by comparing the theoretical milliamp

output of the device under test based on the measured pres-

sure reading, against the actual measured output of the device

under test. The HHC will indicate whether this error is positive

or negative as compared to the theoretical output.

If 50.00 psi is applied to a 0-100 psi pressure transducer with a

1-5 Vdc output, the theoretical output will be 3.000v. If the mea-

sured output from the transducer is 3.015v, the HHC will display

the following:

% Error terms
Input
refers to any pressure or electrical signal input to both the

device under test and the pressure or temperature module in the

HHC.
Input Channel refers either to a pressure or temperature module

installed in the base unit (CH1 or CH2), or a voltage or current

signal connected to the common and input jacks.
Input Range refers to the numeric full scale range of the input

parameter. It is not necessary to program an engineering unit for

the input range. For example, a pressure transducer with a mea-

surement range of 0 to100 psi. The engineering unit is automati-

cally determined by the channel selected (pressure, temperature,

voltage or current).

SECTION 15 h.

PERCENT (%) ERROR FUNCTION

50.000

In(PSi)

Out(volts)

% error

2.423

0.38

Advertising