SECTION 14 l.

MIN/MAX FUNCTION

0.11

inH2O

PSi

15.29

0.033

mA

Loop supply reset failed

–0.12

in H 2 0

-0.012

PSi

SECTION 14 k.

24 VDC POWER SUPPLY ENABLE

AND DISABLE

Remove the excessive load from the circuit and attempt to activate

the 24v loop power supply via the MENU as described above.
If the excessive load is not removed from the circuit upon activa-

tion, the loop supply will be disabled and the following warning

message will be displayed

Note: A direct short between 24Vdc out and COM terminals will

cause complete shut down of the base unit. Remove shorted leads

and power the HHC on with the power key. Base unit will not be

damaged, however warning messages will not be displayed.
If the user correctly removes the excessive load and resets the

loop supply successfully via the MENU 24v loop supply indicator

turns back on and normal function is restored.

Display the minimum and maximum pressure or temperature

readings measured by the installed modules. This function can

be performed once the HHC has been powered on, module “hot

swapped” into the HHC, or values that were zeroed.
Min /Max function is actively updating the display continuously

once it is activated.

To enter Min/Max mode
Press Min/Max key
Press PORT SELECT key to select Channel 1, Channel 2, or

both modules.
Press the ENTER key

0.10

inH2O

PSi

15.34

0.000

mA

Loop supply overloaded