Ashcroft ATE-2 - Handheld Calibrator User Manual
Page 9
7
0.05
4
PS
I
0.86
in
H2
0
The ATE-2 is a simple, yet sophisticated, high accuracy, battery
or USB powered, handheld calibrator device. Used together with
interchangeable pressure and temperature modules, this device
measures, displays and can record pressure, temperature, DC
voltage or milliamp current. Up to three measurements can be
simultaneously displayed.
A standard system consists of a base unit and one or two
Quick-Select modules.
Quick-Select pressure modules are available in over 50 standard
pressure ranges in gauge, absolute, vacuum and compound
ranges capable of measuring pressures from 0.25 inches water
column up to 10,000 pounds per square inch.
Quick-Select temperature modules are designed to work with
several common RTD types including Pt1000, Pt100, Ni120,
Cu10, and up to six programmable resistance curves. Thermo-
couple temperature modules work with Types J, K, T, E, R, S, B
and N thermocouples.
SECTION 2.0
INTRODUCTION