Rtd module – Ashcroft ATE-2 - Handheld Calibrator User Manual

Page 46

Advertising
background image

44

Use DOWN or UP ARROW to highlight “Channel”…

Press ENTER

Use DOWN or UP ARROW to highlight “CH1” “CH2”

Press ENTER
Note: If the incorrect type of module is installed a “MODULE

ERROR” message will be shown.
Return to RTD Module screen

Use DOWN or UP ARROW to highlight “Probe connection…”

Press ENTER

Use DOWN or UP ARROW to highlight “2-wire…”, “3-wire…”, or

“4-wire…”

Press ENTER

Return to RTD Module screen

Use DOWN or UP ARROW to highlight “Probe select…”

Press ENTER

Use DOWN or UP ARROW to highlight “Probe 01 through 08…”

(per table above)

Press ENTER

Return to RTD Module screen

Use DOWN or UP ARROW to highlight “Resolution…”

Press ENTER

Use DOWN or UP ARROW to highlight “1”, “0.1”, ”0.01”or “0.001”

Press ENTER

Return to RTD Module screen

Use DOWN or UP ARROW to highlight “Apply”

Press ENTER
Programmed settings are saved to memory and HHC returns to

Measurement Mode with RTD operating as programmed above.

Note: The maximum resolution that the combination of the HHC

base unit and the RTD interface module and probe can support

can be determined as follows:
The resolution of the RTD module is 0.001% of the full scale

resistance of the reference resistor used in the module. The XXX-

RT1 incorporates a 400 ohm reference resistor and the XXX-RT2

incorporates a 4000 ohm resistor.
Using the XXX-RT1 with a PT100 probe as the example the

maximum resolution is:
0.001% of 400ohms = 0.004 ohms
0.004 ohms/(100 ohms x 0.00385 ohms/degree) =0.0104

degrees C
Therefore rounding to the nearest digit results in a resolution of

SECTION 15 j.

RTD TEMPERATURE

MEASUREMENT CONT.

RTD Module

Channel…

Probe connection…

Probe Select…

Resolution…

Apply

–0.017

PSi

degF

77.9