80

SECTION 16 e.

SYSTEM ACCURACY WITH

THERMOCOUPLE INTERFACE

TABLE I

System Accuracy (reading in temperature units): Includes the HHC base unit and XXX interface module.

(Does not include inaccuracy of the thermocouple device. Consult thermocouple manufacturer or ISA MC96.1

for thermocouple accuracy specifications. Typical inaccuracies range from ±1 to ±2.2°C.)

Max. additional error

Accuracy @ 25°C

Accuracy @ 25°C

due to ambient

Thermocouple

Measurement Range

Ambient (Not including

Ambient (including

temperature deviation

Type

(°C)

internal Reference

internal Reference

from 25°C. Expressed

Junction) Expressed as ±°C

Junction) Expressed as ±°C

as additional C error per

each C deviation from 25°C

J

–210 to –151

0.7

1.1

0.02

–150 to 1200

0.3

0.4

0.01

K

–240 to –201

1.5

2.2

0.05

–200 to –101

1.0

1.5

0.03

–100 to 999

0.5

0.8

0.02

T

–250 to –201

1.5

2.2

0.05

–200 to –101

0.8

1.2

0.03

–100 to 400

0.5

0.8

0.02

E

–250 to –201

1.2

2.0

0.04

–200 to –101

0.6

1.1

0.02

–100 to 400

0.3

0.6

0.01

R&S

–50 to 299

3.4

3.6

0.10

300 to 1768

1.2

1.3

0.04

B

100 to 199

14.0

0

14.0

0

0.44

200 to 499

5.0

5.0

0.16

500 to 999

3.0

3.0

0.08

1000 to 1820

1.2

1.2

0.04

N

–250 to –226

4.3

5.7

0.14

–225 to –101

2.1

2.8

0.07

–100 to 1300

1.0

1.4

0.02

Conversion Factors (to convert °C specifications to other units of measure):

To convert from °C to °F: °F = (1.8 x °C) + 32

To convert from °C to °K: °K = °C + 273.15
To convert from °C to °R: °R = (1.8 x °C) + 427.67

TABLE II

System Accuracy (based on direct millivolt reading from thermocouple – includes HHC base unit and

XXX interface module. (Does not include inaccuracy due to the thermocouple device. Reference junction

not applicable to direct millivolt readings.)

input Range of

Accuracy @ 25°C

Max. additional error due to ambient

temperature deviation from 25°C.

Expressed as additional millivolt error

per each °C deviation from 25°C

–10 to 100 millivolt

± 0.0125 millivolts

.001

Max. additional error

due to ambient

temperature deviation

from 25°C. Expressed

as additional C error per

each C deviation from 25°C

Sandisk 8GB Class 4 SDHC card, SKU SDSDB-8192-A11 .
Other SD cards will work, provided they meet the following

requirements:
1. 4GB to 32 GB SDHC card –or- 4GB to 32GB micro SD card

with micro SD-to-SD adapter

2. Formatted to FAT32 or exFAT filesystem
3. Speed class 4 or higher
4. Write protection disabled

SECTION 16 f.

APPROVED SD DATA CARDS