43

Both modules can display temperature units of Fahrenheit,

Celsius, Kelvin, Rankine, and also support reading output in units

of ohms.
The electrical interface of the probe to the module incorporates

a Switchcraft TA4M type connector and requires an RTD with the

appropriate mating connector. The connector is available through

the supplier of your choice.
Setting up the PTE-2 for an RTD Temperature Measurement
Insert the appropriate XXX-RT Quick-Select RTD Interface

Module.
Connect the desired RTD probe to the Switchcraft connector on

the XXX-RT module.

Note: RTD probes can be connected using two, three or four wire

configurations. Use of four wire configuration is recommended for

optimum accuracy. RTD probes provided by this manufacturer are

of the four wire configuration.
To select RTD probe characteristics
Install an RTD Quick-Select Module and Probe in the HHC

Base Unit.

Press MENU key

Use DOWN or UP ARROW to highlight “Setup…”

Press ENTER

Use DOWN or UP ARROW to highlight “Module…”

Press ENTER

Use DOWN or UP ARROW to highlight “RTD…”

Press ENTER

RTD Module programming screen will appear

XXX-RT1

XXX-RT2

Ohms Measurement

Range

0 - 400 ohms

0 - 4000 ohms

Factory Programmed

RTD Probe 01

Pt100 385: -200 to

500C ±0.15C

Pt1000: 184 to 275C

±0.15C

Factory Programmed

RTD Probe 02

Pt100 392: 550 to

850C ±0.2C

----

Factory Programmed

RTD Probe 03

Ni 120: 80 to 260C

± 0.1C

----

Factory Programmed

RTD Probe 04

Cu 10: 70 to 150C

± 0.6C

----

Available User

Programmable R TD

Probes

4

7

Ohms

+0.01% reading

+0.02ohms

+0.01% reading

+0.2ohms

SECTION 15 j.

RTD TEMPERATURE

MEASUREMENT CONT.