Ashcroft ATE-2 - Handheld Calibrator User Manual

Page 7

Advertising
background image

5

Warning to avoid shock or personal injury

• Follow all applicable equipment safety procedures.
• Inspect the calibrator before use. Do not use the calibrator if

it appears damaged.

• Close battery door and secure with lock screw before using

the calibrator.

• Check all test leads for continuity, damaged insulation or

exposed metal. Replace any damaged leads.

• Use the proper terminals, mode and range for any measur-

ing and sourcing application.

• Connect the common lead before connecting the live test

leads.

• To prevent damage to the unit under test, be sure that the

calibrator is in the correct mode before connecting leads.

• Never apply more than 33 volts between input terminals or

between any terminal and ground.

• When using probes keep fingers away from probe contacts.
• Do not allow water or other liquids inside the calibrator case.
• Do not operate the calibrator near explosive gas or vapor

unless the calibrator is marked as approved for hazardous

area service.

• Never open the ATE-2. There are no field serviceable

components. This will void the warranty.

• When measuring pressure, be sure that the process pres-

sure line is shut off and depressurized before connecting or

disconnecting a pressure module.

• Always use proper tools to avoid damage to pressure

measurement modules. Use a wrench to hold the module

manifold, and a second tool to tighten the mating fitting. See

drawing figure in section 12.

• Replace batteries only with approved AA alkaline batteries

(Quantity 4). Never mix new and used batteries in the cali-

brator.

Warnings for Intrinsically Safe approved versions only
(Refer to Supplement I.S. Manual I&M002-10212)

• When marked as approved the ATE-2 calibrator is agency

approved for use in areas when potentially flammable or

explosive gas or vapor may occur. These areas are referred

to as hazardous (classified) locations in the United States,

as hazardous locations in Canada, as potentially explosive

atmospheres in Europe and as explosive gas atmospheres

in other parts of the world. The HHC’s voltage and current

input jacks have entity parameters and as such can be uti-

lized in a hazardous location to connect to other apparatus

as long as said apparatus meets the requirements of the

entity parameters.

• When marked with the following Symbols the calibrator is

approved by Factory Mutual (FM) for use in Class 1 Division

1, Groups A,B, C,D hazardous areas, by ATEX for use in Ex

ia ll C T4 Ga -20C<Ta<+50C and by CSA for use in Class 1,

Division 1, Groups A, B, C, D

ADDITIONAL WARNINGS

WARNINGS CONT.

Ꮨƽ

Ꮨƽ

Advertising