Pres mod ch1 cal, Enter mod zero – Ashcroft ATE-2 - Handheld Calibrator User Manual

Page 70

Advertising
background image

68

NOTE: Cal Zero function must be performed prior to proceed-

ing to Cal Full Scale; Cal Zero & Cal Full Scale functions must

be performed prior to proceeding to Cal Linearity. The “Cal Full

Scale” & the “Cal Linearity” keys are activated only when this

sequence is observed. The “Apply New Cal” function can be

performed after each individual calibration steps (Zero, Full

Scale, Linearity) was completed.

4. Select “Cal Zero” and Press “ENTER”,
a.

The display will show target Zero pressure (00.00000).

Note: Min and Max allowable limits for all calibration pres-

sure values will be displayed below the target cardinal value.

If the pressure standard does not achieve exactly the target

cardinal pressure value, actual target pressure can be

edited within the Min and Max limits. This value is the pres-

sure reading which is displayed on the external pressure

calibration standard.

The following is defined as “Zero Pressure” for various

pressure types:

Absolute

Absolute Zero (using vacuum pump)

Gauge

Open to atmosphere (vent)

Compound Open to atmosphere (vent)

Differential Open both ports to atmosphere (vent)
Vacuum

Open to atmosphere (vent)

If the cardinal zero value cannot be achieved exactly by the

pressure standard:

b. Press left and RIGHT ARROWs to position cursor beneath

desired character

c. Press UP and DOWN ARROWs to increment numeric

characters

d. When the pressure value displayed on HHC matches value

displayed on pressure standard press ENTER

The display will read “Acquiring Cal Data…Wait while system

is busy”. After Zero Cal is complete (several seconds) the

screen returns to the Calibration Menu with “Cal Full Scale”

selected. At this time the operator can proceed with “Cal Full

Scale”, return to “Zero Cal” to repeat application of zero pres-

sure or save the Zero Cal by selecting “Apply new cal” and

pressing “ENTER”.

5. Select “Cal Full Scale” and press “ENTER”.

Pres Mod CH1 Cal

Cal Zero
Cal Full Scale

Cal Linearity

Apply new cal

Enter Mod Zero

00.00

^

Min -4.00 inH2O

Max 4.00 inH2O

Press Enter to cal

Ꮨƽ

Ꮨƽ

SECTION 15 o.

PRESSURE MODULE CALIBRATION

CONT.