Cal password – Ashcroft ATE-2 - Handheld Calibrator User Manual

Page 69

Advertising
background image

67

0.025% - 0.07% standard accuracy

11 points (approximately 10% increment)

0.1% standard accuracy

6 points (approximately 20% increment

0.25% standard accuracy

5 points (approximately 25% increment)

When calibrating linearity, the HHC will automatically choose the

number and pressure value of data points based on the module’s

full scale pressure range and it’s base accuracy specification.
Pressure module recalibration is a firmware based, menu driven

procedure that provides prompts on the HHC display directing

the operator to apply precision pressure at zero, full scale and

several cardinal points, the quantity of which is determined by

the accuracy specification of the module. Module recalibration is

accessed via the menu system.
NOTE Following Pressure Module Calibration, damping is auto-

matically set to 16 (maximum.). To reduce the damping level to

1 (minimum), power cycle the HHC, or remove and reinstall the

pressure module.
REQUIRED EQUIPMENT
A Precision Pressure standard must be used to maintain the

stated accuracy of the pressure module. It is recommended to

have the pressure standard accuracy 4 times better, than the

accuracy of the calibrated module.
PRESSURE MODULE FIELD CALIBRATION PROCEDURE
1. Insert Pressure Module to be calibrated into the CH 1 Module

bay (left bay, when the HHC faces the operator with module

bays up).

2. Turn the base unit on by pressing the green “ON/OFF” button

on the keypad

3. Initiate the calibration process as follows:
a. Press MENU key

b. Press DOWN or UP ARROW key to highlight “Setup…”

c. Press ENTER

d. Press DOWN or UP ARROW key to highlight “Calibrate”

e.

Press ENTER

Cal Password screen appears. Enter Calibration Password

f.

Press left and RIGHT ARROWs to position cursor beneath

desired character

g.

Press UP and DOWN ARROWs to increment numeric

characters

h. When the password is correct press ENTER
i. “Calibrate” screen appears

Note: If the incorrect password is entered, the HHC display

will show a message “Bad Cal PW”, and the user will be

asked to Retry or Abort.

j.

Press DOWN or UP ARROW key to highlight “Pressure

Module…”

k. Press ENTER
l. “Pres Mod CH1 Cal” screen appears

SECTION 15 o.

PRESSURE MODULE CALIBRATION

CONT.

Cal Password

00000

^

Enter to accept

Esc to cancel