58

The HHC will appear as a Communications Port on a Windows

PC. It will be assigned the port number of the first available port

on the PC. To determine which port number has been assigned

look in the Windows Device Manager. This can be accessed by

the Windows Control Panel under System or by right clicking on

your My Computer Icon, selecting Properties and Device Man-

ager.

In Device Manager expand Ports to see which USB Serial Port is

assigned to the HHC. In the example below the HHC is shown as

COM 8. You may connect more than one HHC to a PC; each will

have its own COM Port assignment.

Terminal or Programming environment setting

The Com Port Selection in the software must be set to the com

port number shown in Device Manager. This Com Port appears

only while the HHC is connected.

Configuration Settings to use:

9600

8 Data Bits

1 stop

No Parity

No handshaking

Note for Terminals: The HHC does not echo characters received,

the user may want to enable local echo.

Modes of Communication:

The HHC can automatically determine which of its three modes is

to be used (Inquiry/Journal, ISO1745 or MODBUS). After Pow-

er-on the HHC uses the first received command from the PC to

establish which mode it will operate in. It will recognize the format

of the first command and remain in that mode until switched by

the user via the Setup Communications Mode MENU or the next

power off/on.

Set up of the HHC Communication modes are available through

the programming MENU.

SECTION 15 m. ii.

Driver Installation Cont.

SECTION 15 m. iii.

Terminal Setup