Delete all, Logs, Clearing – Ashcroft ATE-2 - Handheld Calibrator User Manual

Page 37: Modes

Advertising
background image

35

SECTION 15 f. vii

Clear Internal Memory Storage

Cont.

Use DOWN or UP ARROW to highlight Yes or No

Press ENTER

Selecting “no” will return to the Datalog MENU

Selecting “yes” will proceed to “Clearing Logs…” screen.

When the operation is complete the HHC will return to

Measurement Mode.

The HHC can be used to test pressure and temperature switches.

Onboard firmware provides the ability to verify switch trip point,

reset point and to calculate deadband of a pressure switch.
The HHC looks for a change of state that produces either 2 volts

or a 2 mA change across the corresponding voltage or current

input terminals.
Prior to initiating this test procedure the required electrical set up

should be configured to provide the necessary signal levels per

the wiring diagrams presented in section 14j.
Note: Switch test requires slow changes in process pressure. See

damping to minimum setting.
Set up switch test
The operator must program the following parameters before

beginning switch testing:
• Measurement Channel (Channel 1, Channel 2)
• Input Signal type (Voltage or current)
• Switch type (Normally open or closed).
• See wiring diagrams in Section 14j

Press MENU key

Use DOWN or UP ARROW to highlight “Modes…”

Press ENTER

Use DOWN or UP ARROW to highlight “Switch testing…”

Press ENTER

Ꮨƽ

Delete all

logs?

Are you sure you want to

clear all internal storage?

Yes

No

Clearing

logs…

Clearing all internal

storage: this may take

a little while.

Modes

Dual module functions…

Percent error…

Switch testing…

Communications…

SECTION 15 g.

SWITCH TESTING FUNCTION