7 menu dialogs, 1 parameters ph/u/t and parameters ph/u/t is, Menu dialogs – Metrohm 914 pH / Conductometer User Manual

Page 48: Parameters ph/u/t and parameters ph/u/t is

Advertising
background image

4.7 Menu dialogs

■■■■■■■■■■■■■■■■■■■■■■

38

■■■■■■■■

pH Meter/Conductometer

4.7

Menu dialogs

4.7.1

Parameters pH/U/T and Parameters pH/U/T IS

The Parameters pH/U/T menu dialog for the parameters Measurement
and Calibration is shown below with the structure and the description.

Measuring parame-
ters

Menu dialog for the Measuring parameters.

Measuring mode

Selection dialog for selecting the measuring mode.

pH
The pH value is output.

U
The potential value is output in mV.

T
The temperature is output in °C.

Sensor name

Selection dialog for selecting a sensor from the sensor list.

In instruments with iTrodes, this is only a display field.

Order number

Display field with the sensor's Order number.

Serial number

Display field with the sensor's Serial number.

Temperature

Editing dialog for the manual entry of the measuring temperature.

Default value: 25.0 °C / input range: –999.9 - +999.9 °C

Does not apply for instruments with iTrodes.

Delta measure-
ment mV

Selection dialog

on: with the input field for the Reference
with default value: 0.0 mV / input range: –1,500.0 - +1,500.0 mV

off: is the default setting

Calibration param.

Menu dialog for the Calibration param..

Temperature

Editing dialog for the manual entry of the calibration temperature.

Default value: 25.0 °C / input range: 0.0 - 99.9 °C

Report

Selection dialog

on

off: is the default setting

Number of buf-
fers

Selection dialog for the Number of buffers that are used for calibration.

Default value: 2 / input range: 1 - 5