Metrohm 914 pH / Conductometer User Manual

Page 61

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

4 Operation

pH Meter/Conductometer

■■■■■■■■

51

Immerse the pH electrode in the second buffer solution and con-

tinue the calibration procedure with the

key.

The second buffer solution is measured.

NOTE

Buffer exchange

If the buffer solution was not exchanged, then the message
912-181 Same buffer will appear.

Exchange the buffer solution and continue the calibration with the

key.

3 Result of the calibration

The result of the calibration is displayed in a diagram.

Finish the calibration with the

key.

(The instrument will automatically change to the main dialog after
30 seconds.)

NOTE

Limit values exceeded

If the calibration data is outside the limits defined as calibration
parameters, a corresponding message will be displayed.

You can then accept this calibration data nevertheless with the

key, or you can reject it with the

key and use the exist-

ing calibration data.