4 Operation

pH Meter/Conductometer

■■■■■■■■

45

on

off: is the default setting

pH

Slope lower limit

Editing dialog for the entry of the lower limit value.

Default value: 95.0 / input range: 1.0 - 999.9%

pH

Slope upper
limit

Editing dialog for the entry of the upper limit value.

Default value: 103.0 / input range: 1.0 - 999.9%

pH

Limit value
pH(0)

Selection dialog for selecting whether the limit value is to be applied.

on

off: is the default setting

pH

pH(0) lower limit

Editing dialog for the entry of the lower limit value.

Default value: 6.40 / input range: 0.00 - 99.99

pH

pH(0) upper
limit

Editing dialog for the entry of the upper limit value.

Default value: 8.00 / input range: 0.00 - 99.99

K

Limit value c

Selection dialog for selecting whether the limit value is to be applied.

on

off: is the default setting

K

c lower limit

Editing dialog for the entry of the lower limit value.

Default value: 0.400 /cm / input range: 0.001 - 500 /cm

K

c upper limit

Editing dialog for the entry of the upper limit value.

Default value: 0.550 /cm / input range: 0.001 - 500 /cm

New sensor

Menu dialog with the individual menu lines for entering a new sensor.

Sensor type

Selection dialog for the sensor type.

pH electrode

Conductivity

Metal electrode

Temp. sensor

Other sensor...

Sensor name

Editing dialog for entering the sensor name.

Order number

Editing dialog for entering the Order number.

Serial number

Editing dialog for entering the Serial number.

pH

Slope

Editing dialog for the entry/modification of the Slope.