2 parameters κ/tds/sal/ρ/t, Parameters κ/tds/sal/ρ/t, 2 parameters κ /tds/sal/ ρ /t – Metrohm 914 pH / Conductometer User Manual

Page 49

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

4 Operation

pH Meter/Conductometer

■■■■■■■■

39

Buffer type

Selection dialog for selecting the buffer type.

Available buffers and their values (see Chapter 7.1, page 66)

If required, the preset values may be adjusted for the Special buffer type.
Default value: 7 / input range: –19.999 - +19.999

4.7.2

Parameters Κ/TDS/Sal/ρ/T

The Parameters Κ/TDS/Sal/ρ/T menu dialog for the parameters Mea-
surement
and Calibration is shown below with the structure and the
description.

Measuring parame-
ters

Menu dialog for the Measuring parameters.

Measuring mode

Selection dialog for selecting the Measuring mode.

Conductivity K
The conductivity of the sample is output.

TDS
(Total Dissolved Solids)

Salinity
The salinity is output.

ρ
The resistance value is output.

T
The temperature is output.

Sensor name

Selection dialog for a sensor from the sensor list.

Order number

Display field for the Order number of the selected sensor.

Serial number

Display field for the Serial number of the selected sensor.

Temperature

Editing dialog for the entry of the measuring temperature.

Default value: 25.0 °C / input range: –999.9 - +999.9 °C

Reference temp.

Editing dialog for the entry of the reference temperature of the calibration
standard.

Default value: 25.0 °C / input range: 0 - 99.9 °C

Temp. compens.

Editing dialog for the entry of the temperature compensation value.

Default value: 2.07%/°C / input range: 0.00 - 9.99%/°C

DIN

TDS factor

Editing dialog for the entry of the factor value for the TDS calculation.

Default value: 0.40 / input range: 0.40 - 1.00 °C

Calibration param.

Menu dialog for the Calibration param..

Advertising
This manual is related to the following products: