6 Troubleshooting

pH Meter/Conductometer

■■■■■■■■

63

Problem

Cause

Remedy

The pH value or the tem-
perature of the measuring
solution is not stable.

Measure under exclusion of air.

Regulate the measuring solution's tempera-
ture.

Conductivity is too low
because of an unsuitable
sensor.

Use a suitable sensor.

Measurement takes place
in an organic solution.

Use a suitable sensor.

Non-Metrohm power sup-
ply unit used for charging
the battery.

Use only the supplied power supply unit
during measurement operation.

The measured value
is evidently wrong.

pH calibration is faulty.

Check/repeat calibration.

Check/replace the buffer.

Check the buffer selection in the settings.

Conductivity calibration is
faulty.

Check/repeat calibration.

Check the value for the standard.

Check the value for the reference tempera-
ture.

Check the value for temperature compen-
sation.

The temperature input is
wrong.

Enter the correct measuring temperature.

The wrong temperature
sensor type is selected.

Check the temperature sensor type (Pt1000
or NTC) and select the correct one, if nec-
essary.

The glass membrane or the
diaphragm is contami-
nated.

Clean the membrane or the diaphragm fol-
lowing the instructions in the correspond-
ing leaflet.

The electrolyte is overaged.

Replace the electrolyte.

The sensor is defective.

Replace the sensor.

The slope is insuffi-
cient during calibra-
tion.

The glass membrane or the
diaphragm is contami-
nated.

Clean the electrode following the instruc-
tions in the leaflet.