4 Operation
pH Meter/Conductometer
■■■■■■■■
47
■
off: is the default setting
K
c lower limit
Editing dialog for the entry of the lower limit value.
■
Default value: 0.400 /cm / input range: 0.001 - 500 /cm
K
c upper limit
Editing dialog for the entry of the upper limit value.
■
Default value: 0.550 /cm / input range: 0.001 - 500 /cm
Delete sensor
Selection dialog for deleting a sensor.
The data will be irreversibly deleted.
4.7.5
Report
The Report menu dialog is shown below with the structure and the
description.
NOTE
Printer
pH Meter/Conductometer instruments support various printer types for
report output. If your printer is not listed, please use the printer
Universal (ESC-POS), which has appropriate setting parameters.
Report
Selection dialog for data output in the Report.
■
Calibration pH
■
Calibration pH IS
■
Calibration κ
■
Sensors
■
Configuration
■
Parameters pH
■
Parameters pH IS
■
Parameters κ
■
Meas. values
■
All reports
Line feed
Editing dialog to indicate the lines to be inserted at the end of the report.
■
Default value: 2 lines / input range: 0 - 99 lines
Printer
Selection dialog to indicate the printer for report output.
■
HP Officejet Pro
Page printer with paper size A4
■
HP Laserjet Pro
Page printer with paper size A4
■
Epson (ESC-POS)
Roll printer with paper width 80 mm