7 merck certipur 25, Merck certipur 25 – Metrohm 914 pH / Conductometer User Manual

Page 83

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

7 Appendix

pH Meter/Conductometer

■■■■■■■■

73

7.1.7

Merck CertiPUR 25

Table 12

Merck CertiPUR 25 buffer solutions

Temp.

(°C)

Merck CertiPUR (25°C)

pH

4.00

pH

7.00

pH

9.00

pH

10.00

0

-

-

-

-

5

4.05

7.09

9.22

10.22

10

4.04

7.08

9.16

10.18

15

4.02

7.04

9.10

10.10

20

4.01

7.02

9.06

10.06

25

4.00

7.00

9.00

10.00

30

3.99

6.98

8.98

9.94

35

3.98

6.98

8.93

9.90

40

3.98

6.97

8.89

9.86

45

3.98

6.97

8.86

9.80

50

3.98

6.97

8.84

9.73

55

-

-

-

-

60

-

-

-

-

65

-

-

-

-

70

-

-

-

-

75

-

-

-

-

80

-

-

-

-

85

-

-

-

-

90

-

-

-

-

95

-

-

-

-

NOTE

Update

The values of the individual buffers with the corresponding tempera-
tures are kept up to date as far as possible. However, they may be
changed by the different manufacturers.