8 beckmann, Beckmann – Metrohm 914 pH / Conductometer User Manual

Page 84

Advertising
background image

7.1 Saved buffer series

■■■■■■■■■■■■■■■■■■■■■■

74

■■■■■■■■

pH Meter/Conductometer

7.1.8

Beckmann

Table 13

Beckmann buffer solutions

Temp.

(°C)

Beckmann

pH

4.00

pH

7.00

pH

10.00

0

4.00

7.12

10.32

5

4.00

7.09

10.25

10

4.00

7.06

10.18

15

4.00

7.04

10.12

20

4.00

7.02

10.06

25

4.00

7.00

10.01

30

4.01

6.99

9.97

35

4.02

6.99

9.93

40

4.03

6.98

9.89

45

4.05

6.98

9.86

50

4.06

6.97

9.83

55

4.08

6.98

-

60

4.09

6.98

-

65

4.11

6.99

-

70

4.12

6.99

-

75

4.14

7.00

-

80

4.16

7.00

-

85

4.18

7.01

-

90

4.19

7.02

-

95

4.21

7.03

-

NOTE

Update

The values of the individual buffers with the corresponding tempera-
tures are kept up to date as far as possible.

However, they may be changed by the respective manufacturers.

Advertising