2 measurement, Measurement – Metrohm 914 pH / Conductometer User Manual

Page 64

Advertising
background image

4.9 Conductivity measurement

■■■■■■■■■■■■■■■■■■■■■■

54

■■■■■■■■

pH Meter/Conductometer

Start the calibration with the

key.

Rinse the conductivity sensor with water and immerse it in the

first standard solution and then confirm with the

key.

The calibration temperature is measured with the connected tem-
perature sensor and added to the calibration data.
If no temperature sensor is connected, then the temperature has
to be entered manually.

Enter the reference temperature for the standard solution.

Enter the conductivity value of the standard solution at the refer-
ence temperature.

Enter the coefficient for the current temperature and the selected
reference temperature for temperature compensation.

Trigger the calibration with the standard solution with the
key.

2 Result of the calibration

The result of the calibration (cell constant) is recorded and stored
for the respective sensor.

Calibration is completed and the instrument will automatically
change back to the main dialog after 30 seconds.

NOTE

Limit values exceeded

If the calibration data is outside the limits defined as calibration
parameters, a corresponding message will be displayed.

You can then accept this calibration data nevertheless with the

key, or you can reject it with the

key.

4.9.2

Measurement

NOTE

Measured value criteria

You can set the various criteria for defining the measured value deter-
mination as follows (see Chapter 4.7.3, page 40):

1 Selecting the printout criterion