4 Operation

pH Meter/Conductometer

■■■■■■■■

57

NOTE

Data output

For a data output, save or print+save has to be set for the recording
of measured values so that this data is saved on the instrument.

Data generation

The measured value data has to be re-generated before any data trans-
fer.

The data on the instrument memory can be saved in two data formats:

PC/LIMS report
Data in PC/LIMS format can be imported into and processed in the
Metrohm program tiBase for evaluation.

CSV file
CSV data can be imported as text data in MS Excel and processed for
evaluation.

The following parameters are required for text conversion:

Data type = separated with a semicolon

Data source = Unicode (UTF-8)

Data format of the rows = standard

Generating/transferring report data

NOTE

USB interface

If the report data is to be generated while the instrument is connected
with the PC, then the connection will be briefly interrupted.

After the report data has been generated, the connection will be auto-
matically established again.

1 Generating report data

You can generate the report data by selecting:

save as PC/LIMS or save as CSV

under Menu

Measured values Values.