3 din (according to din standard 19267, 2012), Din (according to din standard 19267, 2012) – Metrohm 914 pH / Conductometer User Manual

Page 79

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

7 Appendix

pH Meter/Conductometer

■■■■■■■■

69

7.1.3

DIN (according to DIN standard 19267, 2012)

Table 8

DIN buffer solutions

Temp.

(°C)

DIN (according to DIN standard 19267, 2012)

pH

1.00

pH

3.00

pH

4.00

pH

7.00

pH

9.00

pH

12.00

0

1.08

-

4.67

6.89

9.48

-

5

1.08

-

4.66

6.86

9.43

-

10

1.09

3.10

4.66

6.84

9.37

13.37

15

1.09

3.08

4.65

6.82

9.32

13.15

20

1.09

3.07

4.65

6.80

9.27

12.96

25

1.09

3.06

4.65

6.79

9.23

12.75

30

1.10

3.05

4.65

6.78

9.18

12.61

35

1.10

3.05

4.66

6.77

9.13

14.44

40

1.10

3.04

4.66

6.76

9.09

12.29

45

1.10

3.04

4.67

6.76

9.04

12.13

50

1.11

3.04

4.68

6.76

9.00

11.98

55

1.11

3.04

4.69

6.76

8.97

11.84

60

1.11

3.04

4.70

6.76

8.92

11.69

65

1.11

3.04

4.71

6.76

8.90

11.56

70

1.11

3.04

4.72

6.76

8.88

11.43

75

1.12

3.04

4.74

6.77

8.86

11.30

80

1.12

3.05

4.75

6.78

8.85

11.19

85

1.12

3.06

4.77

6.79

8.83

11.08

90

1.13

3.07

4.79

6.80

8.82

10.99

95

-

-

-

-

-

-

NOTE

Update

The values of the individual buffers with the corresponding tempera-
tures are kept up to date as far as possible.

However, they may be changed by the respective manufacturers.

Advertising
This manual is related to the following products: