4 fisher, Fisher – Metrohm 914 pH / Conductometer User Manual

Page 80

Advertising
background image

7.1 Saved buffer series

■■■■■■■■■■■■■■■■■■■■■■

70

■■■■■■■■

pH Meter/Conductometer

7.1.4

Fisher

Table 9

Fisher buffer solutions

Temp.

(°C)

Fisher

pH

2.00

pH

4.00

pH

7.00

pH

10.00

0

-

4.01

7.13

10.34

5

1.98

3.99

7.10

10.26

10

1.98

4.00

7.07

10.19

15

2.02

3.99

7.05

10.12

20

2.00

4.00

7.02

10.06

25

2.00

4.00

7.00

10.00

30

2.00

4.01

6.99

9.94

35

2.02

4.02

6.98

9.90

40

2.01

4.03

6.97

9.85

45

2.01

4.04

6.97

9.81

50

2.01

4.06

6.97

9.78

55

-

4.07

6.97

9.74

60

-

4.09

6.98

9.70

65

-

4.11

6.99

9.68

70

-

4.13

7.00

9.65

75

-

4.14

7.02

9.63

80

-

4.16

7.03

9.62

85

-

4.18

7.06

9.61

90

-

4.21

7.08

9.60

95

-

4.23

7.11

9.60

NOTE

Update

The values of the individual buffers with the corresponding tempera-
tures are kept up to date as far as possible.

However, they may be changed by the respective manufacturers.

Advertising
This manual is related to the following products: