4 Operation

pH Meter/Conductometer

■■■■■■■■

41

print:

save:

print+save:

Criterion

Selection dialog to indicate when the Measured values are applied during the
measurement.

immediately
The displayed measured value will be immediately applied.

time-dependent
The measured value will be applied during a Time interval that can be set.
For the PC/LIMS report, the individual measured values are grouped in a data
group.
The interval ends after the Stop time has been achieved.
The parameters are as follows:

Time interval in seconds

Default value: 4 s / input range: 1 - 999,999 s

Stop time in seconds

Default value: off / input range: 1 - 999,999 s

when changed
The subsequent measured value will only be applied automatically if the devi-
ation from the previous measured value is greater than the value
delta pH/T/mV/Κ defined here.

– Delta pH

Default value: 0.50 pH / input range: 0.10 - 16.00 pH

– Delta T(pH)

Default value: 0.5 °C / input range: 0.1 - 100.0 °C

– Delta mV

Default value: 30.0 mV / input range: 0.1 - 999.9 mV

– Delta κ

Default value: 0.1 mS/cm / input range: 0.0001 - 10 mS/cm

– Delta T(κ)

Default value: 0.5 °C / input range: 0.1 - 100.0 °C

– Stop time

Default value: off / input range: 1 - 999,999 s

– Primary channel

Selection dialog for selecting the measuring channel that has to fulfill
the change criterion.