9 conductivity measurement, 1 determination of the cell constant (calibration), Conductivity measurement – Metrohm 914 pH / Conductometer User Manual

Page 63: Determination of the cell constant (calibration)

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

4 Operation

pH Meter/Conductometer

■■■■■■■■

53

NOTE

Measured value recording

Depending on the settings of the measured value criterion, the
recording of the measured value may take some time. During the
measurement, hold the sensor steady and do not touch the sam-
ple vessel with it.

For measurements that take longer, we recommend using a stand
to secure the sensor in place.

Finishing measurements

After the last measurement, rinse the sensor and follow the stor-
age instructions for the sensor.

4.9

Conductivity measurement

This chapter describes the required steps to carry out a simple conductivity
measurement with calibration. The description is limited to only the indis-
pensable steps and will enable you to carry out first measurements with
the instrument directly.

4.9.1

Determination of the cell constant (calibration)

NOTE

Measuring channel selection

In order to perform the calibration, you have to select the correspond-
ing measuring channel in the main dialog.

You cannot perform a calibration in the two-channel view in the main
dialog.

Determination of the cell constant

The calibration parameters are set to default values (see Chapter 4.7.2,
page 39)
.

If the Report selection dialog is set to on in the Calibration param.
menu dialog, then the calibration data will be output immediately.

1 Starting the calibration