9 radiometer, Radiometer – Metrohm 914 pH / Conductometer User Manual

Page 85

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

7 Appendix

pH Meter/Conductometer

■■■■■■■■

75

7.1.9

Radiometer

Table 14

Radiometer buffer solutions

Temp.

(°C)

Radiometer

pH

4.010

pH

7.000

pH

9.180

0

4.000

7.118

9.464

5

3.998

7.087

9.395

10

3.997

7.059

9.332

15

3.998

7.036

9.276

20

4.001

7.016

9.225

25

4.005

7.000

9.180

30

4.011

6.987

9.139

35

4.018

6.977

9.102

40

4.027

6.970

9.068

45

4.038

6.965

9.038

50

4.050

6.964

9.011

55

4.064

6.965

8.985

60

4.080

6.968

8.962

65

4.097

6.974

8.941

70

4.116

6.982

8.921

75

4.137

6.992

8.900

80

4.159

7.004

8.885

85

4.183

7.018

8.867

90

4.210

7.034

8.850

95

4.240

-

-

NOTE

Update

The values of the individual buffers with the corresponding tempera-
tures are kept up to date as far as possible.

However, they may be changed by the respective manufacturers.

Advertising