7.1 Saved buffer series

■■■■■■■■■■■■■■■■■■■■■■

68

■■■■■■■■

pH Meter/Conductometer

7.1.2

NIST (according to DIN standard 19266, 2000)

Table 7

NIST buffer solutions

Temp.

(°C)

NIST (according to DIN standard 19266, 2000)

pH

1.000

pH

4.000

pH

7.000

pH

9.000

pH

13.000

0

-

4.010

6.984

9.464

13.423

5

1.668

4.004

6.950

9.392

13.207

10

1.670

4.001

6.922

9.331

13.003

15

1.672

4.001

6.900

9.277

12.810

20

1.676

4.003

6.880

9.228

12.627

25

1.680

4.008

6.865

9.184

12.454

30

1.685

4.015

6.853

9.144

12.289

35

1.691

4.025

6.843

9.107

12.133

40

1.697

4.036

6.837

9.076

11.984

45

1.704

4.049

6.834

9.046

11.841

50

1.712

4.064

6.833

9.018

11.705

55

1.715

4.075

6.834

8.985

11.574

60

1.723

4.091

6.836

8.962

11.449

65

1.732

4.108

6.840

8.941

-

70

1.743

4.126

6.845

8.921

-

75

1.754

4.145

6.852

8.902

-

80

1.766

4.164

6.859

8.885

-

85

1.778

4.185

6.867

8.867

-

90

1.792

4.205

6.877

8.850

-

95

1.806

4.227

6.886

8.833

-

NOTE

Update

The values of the individual buffers with the corresponding tempera-
tures are kept up to date as far as possible.

However, they may be changed by the respective manufacturers.