7.1 Saved buffer series

■■■■■■■■■■■■■■■■■■■■■■

72

■■■■■■■■

pH Meter/Conductometer

7.1.6

Merck CertiPUR 20 / Titrisol

Table 11

Merck CertiPUR 20 / Titrisol buffer solutions

Temp.

(°C)

Merck CertiPUR 20

pH

2.000

pH

4.000

pH

7.000

pH

9.000

pH

12.000

0

2.010

4.050

7.130

9.240

12.580

5

2.010

4.040

7.070

9.160

12.410

10

2.010

4.020

7.050

9.110

12.260

15

2.000

4.010

7.020

9.050

12.100

20

2.000

4.000

7.000

9.000

12.000

25

2.000

4.010

6.980

8.950

11.880

30

2.000

4.010

6.980

8.910

11.720

35

2.000

4.010

6.960

8.880

11.670

40

2.000

4.010

6.950

8.850

11.540

45

2.000

4.000

6.950

8.820

11.440

50

2.000

4.000

6.950

8.790

11.330

55

2.000

4.000

6.950

8.760

11.190

60

2.000

4.000

6.960

8.730

11.040

65

2.000

4.000

6.960

8.715

10.970

70

2.010

4.000

6.960

8.700

10.900

75

2.010

4.000

6.960

8.680

10.800

80

2.010

4.000

6.970

8.660

10.700

85

2.010

4.000

6.980

8.650

10.590

90

2.010

4.000

7.000

8.640

10.480

95

-

4.000

7.020

-

-

NOTE

Update

The values of the individual buffers with the corresponding tempera-
tures are kept up to date as far as possible. However, they may be
changed by the different manufacturers.

The CertiPUR 20 and Titrisol product lines have the identical values.