6.3 Restarting/resetting the instrument

■■■■■■■■■■■■■■■■■■■■■■

64

■■■■■■■■

pH Meter/Conductometer

Problem

Cause

Remedy

No hydrated layer is pres-
ent on the glass membrane
after measurements in
water-free solutions.

Hydrate the electrode between the meas-
urements.

The buffer solutions are not
OK.

Replace the buffer solutions.

The sensor is "worn out".

Replace the sensor.

6.3

Restarting/resetting the instrument

6.3.1

Instrument reset

In case of a malfunction, the instrument might not work correctly any-
more and not be switched off.

You can switch off the instrument using the following key combination
for approx. two seconds:

The instrument can be switched on again.

NOTE

Data storage

The currently measured data and modified settings cannot be saved if
the instrument is reset.

6.3.2

Resetting the instrument to factory settings

This function deletes all user data on the instrument. Afterwards, the
instrument will be in the state as delivered from the manufacturer with the
default settings.

CAUTION

User data

The user data will be irreversibly deleted.

You can reset the instrument with the following key combination during
startup: