4 sensors, Sensors – Metrohm 914 pH / Conductometer User Manual

Page 52

Advertising
background image

4.7 Menu dialogs

■■■■■■■■■■■■■■■■■■■■■■

42

■■■■■■■■

pH Meter/Conductometer

drift-dependent
The measured value will be applied when the value is stable according to the
drift criterion.
The drift thresholds are preset and cannot be changed:

– pH measurement: 0.028 pH/min
– Potential measurement U/mV: 1.875 mV/min
– Temperature measurement T/°C: 0.974 °C/min

For the conductivity, various drift thresholds are stored (depending on the
measuring range):

– 0.005 mS/cm/min up to 16 µS/cm
– 0.5 mS/cm/min between 16 µS/cm and 1 mS/cm
– 10 mS/cm/min for 1 mS/cm and higher

For instruments with two measuring channels, the primary measuring chan-
nel for the fulfillment of the drift criterion has to be selected.

– Primary channel

Selection dialog for selecting the measuring channel that has to fulfill
the drift criterion.
pH/mV or Cond.

Output date/time

Selection dialog for selecting whether a time stamp is to be assigned to a mea-
sured value.

on: Measured values contain a time stamp on the report.

off: Measured values do not contain a time stamp on the report.
Default value: off

Output headers

Selection dialog for selecting how the headers are output.

once

always: is the default setting

off

Calibration data

Selection dialog for selecting whether the main Calibration data is assigned for
the output of the measured values.

on

off: is the default setting

4.7.4

Sensors

The Sensors menu dialog is shown below with the structure and the
description.