8 Technical specifications

pH Meter/Conductometer

■■■■■■■■

79

8 Technical specifications

8.1

Measuring inputs

The four available instruments are each equipped with specific measuring
inputs.

The following table lists the measuring inputs for each instrument and the
corresponding measuring modes.

Table 18

Measuring inputs / instrument

Device

Measuring inputs / measuring modes

Electrode,

analog

Electrode,

digital

Conductivity

Tempera-

ture

Reference

2.912.010

Κ/TDS/Sal/ρ/T

2.913.010

pH/U/T

pH/U/T

T

x

2.914.010

pH/U/T

Κ/TDS/Sal/ρ/T

2.914.020

pH/U/T

Κ/TDS/Sal/ρ/T

T

x

Table 19

Specification of the measuring inputs

Measuring range

Resolution

Measuring accuracy

1)

pH value

–13.000 - +20.000

0.001 pH

±0.003 pH

Temperature:

Pt1000

with iCon-
nect

NTC 30 kΩ

–150 °C - +250 °C

–150 °C - +250 °C

–5 °C - +250 °C

0.1 °C

0.1 °C

0.1 °C

±0.2 °C (–20 °C - +150 °C)

±0.4 °C (–20 °C - +150 °C)

±0.6 °C (+10 °C - +40 °C)

Potential

–1,200.0 mV - +1,200.0 mV

0.1 mV

±0.2 mV

Conductance

2)

0.1 µS - 500 mS

4 significant
places

±0.5% at 0.1 µS - 16 µS

±0.5% at 16 µS - 1 mS

±1.0% at 1 mS - 500 mS

1)

±1 digit, without sensor error, at reference conditions

2)

To obtain the conductivity, the corresponding value must be multiplied

by the cell constant. The indicated values apply for c = 1/cm.

Display interval of the measurement = 1 s