CANON Digital Galvano Scanner System GM-1000 Series
Users Manual 1.20

119

Parameter ID

28

Parameter

Name

Scan Angle of Raster Scan

Data

Raster scan angle (Unit: degree × 10000)

Return Value 0: Parameter setting successful

1: Parameter setting unsuccessful

Explanation

This parameter is used to set the angle of raster scan (operation by command).
(See Explanation of Parameter ID = 26.)
Set the scan angle × 10000.
For ±5°, set 50000 (5 × 10000).

Related
Command
/Parameter

Cycle Time of Raster Scan (Parameter ID = 26)
Duty of Raster Scan (Parameter ID = 27)
Interval Time of Raster Scan (Parameter ID = 30)
Start Position of Raster Scan (Parameter ID = 31)

Parameter ID

30

Parameter

Name

Interval Time of Raster Scan

Data

Raster scan Interval time (Unit: second × 100)

Return Value 0: Parameter setting successful

1: Parameter setting unsuccessful

Explanation

This parameter is used to set the Interval time of raster scan (operation by command).
(See Explanation of Parameter ID = 26.)
Set the wait time (sec) × 100.
For 0.1 sec, set 10 (0.1 × 100).


Related
Command
/Parameter

Cycle Time of Raster Scan (Parameter ID = 26)
Duty of Raster Scan (Parameter ID = 27)
Scan Angle of Raster Scan (Parameter ID = 28)
Start Position of Raster Scan (Parameter ID = 31)





CW

CCW

Raster Scan Interval Time

0

Scan Angle 
(-)

Raster Scan Interval Time

Scan Angle 
(+)

CW

CCW

Raster Scan Interval Time

0

Scan Angle 
(-)

Raster Scan Interval Time

Scan Angle 
(+)