X and y axis matching – Canon GM-1000 User Manual

Page 87

Advertising
background image

CANON Digital Galvano Scanner System GM-1000 Series
Users Manual 1.20

87

7.4. X and Y Axis Matching

It is possible to match the step response waveforms of the X axis (Axis #1) and Y axis (Axis #2).

(Summary of Operations)

• Matching the frequency response

This operation matches the frequency response of both axes at 1 kHz. Gain is adjusted with “loop

gain fine adjustment” (parameter ID = 45). Phase is adjusted with “Notch filter Q value” (parameter

ID = 17 and 20) of each axis.

• Matching the step response overshoot

This operation matches the overshoot of the step response waveforms for each axis. It is carried

out after matching the frequency response. Overshoot is adjusted with “Over-shoot

Control“(Parameter ID = 42).

(Operating Conditions)

• Both axis motors are connected to controller.
• Both axis use digital notch filter for tuning.

The X and Y match may not function in some situations other than the above conditions. In this case

an error will occur during tuning, and it will stop.

Advertising