Rice Lake Counterpart Configurable Counting Indicator - User Manual Rev. E - Version 2.5 User Manual
Page 106
Advertising
100
Counterpart Counting Scale Indicator
3. Install jumpers on JP5 and JP6. See Table 2-5 on page 16.
4. Plug in power to the Counterpart and press the indicator’s
Power
button.
5. With Revolution open, begin a new configuration file for the Counterpart.
6. Click
Update
Counterpart
Firmware
.
Figure 9-4. Revolution Screen
7. The Rice Lake Updater screen appears. Specify the COM port the Counterpart is connected to, and click
the ellipses (...) to browse to and select the desired .hex file.
Figure 9-5. Rice Lake Updater Screen
8. Click the
Program
button. The update will take several moments.
9. When complete, remove the jumpers shown in Figure 2-13 on page 15 and press the
Power
button to power
up the indicator.