74

Counterpart Counting Scale Indicator

6.1.4

Clear and Reset Commands

The following commands can be used to clear and reset the Counterpart:

RS

: Reset system. Resets the indicator without resetting the configuration.

RESETCONFIGURATION

: Restores all configuration parameters to their default values (menu mode only). The

RESETCONFIGURATION

function can also be initiated by pressing navigating to the

DEFALT

parameter under the

VERS menu and selecting YES. Then press

ENTER

to reset the indicator.

Note

All load cell calibration settings are lost when the RESETCONFIGURATION command is run.

CLEARBUFFER

: Clears the Weighvault data buffer.

6.1.5

Parameter Setting Commands

Parameter setting commands allow you to display or change the current value for a particular configuration
parameter.

Current configuration parameter settings can be displayed in all modes using the following syntax:

command<ENTER>

Most parameter values can be changed in menu mode only.

Use the following command syntax when changing parameter values:

command=value<ENTER>

, where

value

is either

a number or a parameter value. Use no spaces before or after the equal (=) sign. If you type an incorrect command,
the display reads

??

.

For example, to set the motion band parameter to 5 divisions, type the following:

SC.MOTBAND=5<ENTER>

For parameters with selectable values, enter the command and equal sign followed by a question mark:

command=?<ENTER>

to see a list of those values. The indicator must be in menu mode to use this function.

Note

Some parameters are valid only if other parameters or parameter values are specified. Restrictions for front-
panel configuration also apply to serial command configuration.

Command

Description

Values

SC.GRADS#n

Graduations

1–100000

SC.ZTRKBND#n

Zero track band

0, 0–100

SC.ZRANGE#n

Zero range

1.900000, 0–100

SC.MOTBAND#n

Motion band

1, 0–100

SC.SSTIME#n

Standstill time

1–65535

SC.OVRLOAD#n

Overload

FS+2%, FS+1D, FS+9D, FS

SC.DIGFLTR1#n
SC.DIGFLTR2#n
SC.DIGFLTR3#n

Digital filtering

1, 2, 4, 8, 16, 32, 64, 128, 256

SC.DFSENS#n

Digital filter cutout sensitivity

2OUT, 4OUT, 8OUT, 16OUT, 32OUT,
64OUT, 128OUT

SC.DFTHRH#n

Digital filter cutout threshold

NONE, 2D, 5D, 10D, 20D, 50D, 100D,
200D, 250D

SC.THRESH#n

Zero threshold

09999999

SC.SMPRAT#n

Sample rate

30HZ, 60HZ, 120HZ, 240HZ, 480HZ,
960HZ

SC.PWRUPMD#n

Power up mode

GO, DELAY

SC.TAREFN#n

Tare function

BOTH, NOTARE, PBTARE, KEYED

SC.PRI.DECPNT#n

Primary units decimal position

8.888888, 88.88888, 888.8888, 8888.888,
88888.88, 888888.8, 8888888, 8888880,
8888800

Table 6-4. Scales Serial Commands