Configuration

31

Scale Calibration Menu

Calibration can be performed in two places within the menu: the CALIBR menu shown in Figure 3-6 on page 26 is
an in-depth scale setup and calibration.

A “quick access” calibration is shown in Figure 4-1 on page 55.

Display and edit

test weight value

Same as POINT 1

Previous A/D raw

counts are shown

Press calibrate

softkey to calibrate

zero. Press menu

to cancel

calibrating.

Previous A/D raw

counts are shown

Press calibrate

softkey to calibrate

zero. Press menu

to cancel

calibrating.

Calibrating, please wait

Press enter to

remove offset from

zero and span

calibrations

POINT 1

POINT 2

POINT 3

POINT 4

POINT 5

WZERO

WVAL

WSPAN

REZERO

WLIN

WGT 1

CAL1

CALIBR

Figure 3-9. Scale Calibration Menu Layout

Table 3-8. Scale Calibration Menu Parameters

Parameter

Choices

Description

WZERO

Allows you to view or change the zero calibration AD count value.

WVAL

Allows you to view and edit the test weight value.

WSPAN

Allows you to view or change the span calibration AD count value.

WLIN

POINT 1-5

Press ENTER to display and edit test weight and calibration values for up to five linearization
points. Perform linear calibration only after WZERO and WSPAN have been set.

Parameters include CAL1 — View or change linear point calibration value.

WGT1 — Sets test value.

REZERO

Removes offset from zero and span calibrations. Can be used during a span calibration
process using hooks and chains.