55

4.0

Calibration

The Counterpart can be calibrated using the front panel, EDP commands, or Revolution

®

.

Note

Calibration can be performed in two places within the menu: the CALIBR menu shown in Figure 4-1 and the
SCALE sub-menu shown in Figure 3-8 on page 28. The CAL
IBR menu shown in Figure 4-1 is a “quick access”
calibration; for more in-depth scale setup and calibration, use the menus found under SETUP

/

CONFIG

/

SCALE (see Figure 3-8 on page 28).

Calibration consists of the following steps:

Zero calibration

Entering the test weight value

Span calibration

Optional five-point linearization

Optional rezero calibration for test weights using hooks or chains.

Note

The

Counterpart

requires the WZERO and WSPAN points to be calibrated. The linearity points are optional,

but must not duplicate zero or span. During calibration, the TARE (Enter) key acts as a data entry confirmation
key. The soft calibrate key accepts that value.

Display and edit

test weight value

Same as POINT 1

Previous A/D raw

counts are shown

Press calibrate

softkey to calibrate

zero. Press menu

to cancel

calibrating.

Previous A/D raw

counts are shown

Press calibrate

softkey to calibrate

zero. Press menu

to cancel

calibrating.

Calibrating, please wait

Press enter to

remove offset from

zero and span

calibrations

POINT 1

POINT 2

POINT 3

POINT 4

POINT 5

WZERO

WVAL

WSPAN

REZERO

WLIN

WGT 1

CAL1

ID

DISPLAY

TEST

SETUP

AUDIT

CALIBR

SETPTS

Figure 4-1. Calibration (CALIBR) Menu

4.1

Front Panel Calibration

1. Press the

MENU SETUP

button, and then press until the display reads

CALIBR

.

2. Press

TARE (Enter)

or to go to the scale to be calibrated. Press

TARE (Enter)

or to go to

WZERO

. Press

TARE (Enter)

again to display a value and press the

Calibrate

softkey

to Calibrate

WZERO.

Note

If you do not want to calibrate zero, press Menu to exit.

3. The indicator displays

CALIBRATING

while calibration is in progress. When complete, press

TARE (Enter)

or

to advanced to the next prompt (

WVAL

).

Note

To view the new AD count, repeat Step 3, but instead of pressing TARE (Enter) while viewing the value, press
MENU SETUP to exit.

4. With

WVAL

displayed, press

TARE (Enter) or

to display or edit the stored calibration weight value. Press

TARE (Enter)

to store the

WVAL

value and advance to

WSPAN

.

5. With

WSPAN

displayed, press

TARE (Enter)

or to view the A/D count. Place test weights on the scale equal

to WVAL. Press the

Calibrate

softkey to calibrate WSPAN.