Note

56

Counterpart Counting Scale Indicator

If you do not want to calibrate span, press Menu to exit.

6. After the

Calibrate

softkey is pressed, the indicator displays

CALIBRATING

when complete. Press

TARE

(Enter)

or and the

Save Exit

softkey or proceed to

WLIN

.

Note

To view the new AD count, navigate back to the WSPAN, and repeat Step 6; however, instead of pressing the
Calibrate softkey while viewing the value, press Save/Exit to exit.

4.2

Five-point Linearization

Five-point linearization (using the WLIN parameter) provides increased scale accuracy by calibrating the indicator
at up to five additional points between the zero and span calibrations.

Linearization is optional: if you choose not to perform linearization, skip the WLIN parameter; if linearization
values have previously been entered, these values are reset to zero during calibration. WZERO and WSPAN must
be calibrated before adding linearization points; WLIN values must be less than the WSPAN value and must not
duplicate WZERO or WSPAN.

To perform linearization, follow the procedure below:

With

WLIN

displayed, press

Down

to go to the first linearization point (

POINT 1

) and then press

Down

again to show

the weight value prompt (

WGT 1

). Then press

Down

to show the weight value. Place test weights on the scale, and

use the numeric keypad to enter the actual test weight value. Press

enter

to save the value and move to the

calibration (

CAL 1

) prompt. Press

down

to show the current calibration value, and then press the

Calibrate

softkey to

calibrate the linearization point. When complete, the AD count for the linear calibration is displayed. Press

enter

again to save the calibration value and go to the next prompt (

POINT 2

).

Repeat for up to five linearization points. To exit the linearization parameters, press the

Up

key to return to WLIN.

4.3

Rezero

The optional rezero function is used to remove a calibration offset when hooks or chains are used to hang the test
weights.

Note

The rezero function cannot be used with five-point linear calibration.

If no other apparatus was used to hang the test weights during calibration, remove the test weights and press

Up

to

return to the CALIBR sub-menu.

If hooks or chains were used during calibration, remove these and the test weights from the scale. With all weight
removed, go to the REZERO parameter, and press

Down

to show the current zero value. Press the

Calibrate

softkey

to adjust the zero and span calibration values. Press

enter

or

Up

to return to the CALIBR sub-menu.

Press

Up

to return to the SCALES menu, or press the

Save and Exit

softkey to exit setup mode.

4.4

EDP Command Calibration

To calibrate the indicator using EDP commands, the indicator EDP port must be connected to a terminal or
personal computer. See Section 2.8.1 on page 12 for EDP port pin assignments.

Once the indicator is connected to the sending device, do the following:

1. Place the indicator in config mode (display must read

SCALE

— see

Figure 3-8 on page 28) and remove all

weight from the scale platform. If your test weights require hooks or chains, place the hooks or chains on
the scale for zero calibration.

2. Send the SC.WZERO#n command to calibrate zero. The indicator displays

CALIBRATING

while calibration

is in progress.

3. Place test weights on the scale and use the SC.WVAL#n command to enter the test weight value in the

following format:

SC.WVAL=nnnnnn<CR>

4. Send the SC.WSPAN#n command to calibrate span. The indicator displays

CALIBRATING

while calibration

is in progress.