6 calibration menu, Figure 3-6, Counterpart – Rice Lake Counterpart Configurable Counting Indicator - User Manual Rev. E - Version 2.5 User Manual

Page 32

Advertising
background image

26

Counterpart Counting Scale Indicator

3.6

Calibration Menu

See Section 4.0 on page 55 for calibration procedures. The Calibration menu can be protected by assigning a
password in the Feature menu.

Note

The

Counterpart

requires the WZERO and WSPAN points to be calibrated. The linearity points are optional,

but must NOT duplicate zero or span. During calibration, the TARE key acts as a data entry confirmation key. It
also acts as an EXECUTE key and accepts the value if calibration was successful.

Display and edit

test weight value

Same as POINT 1

Previous A/D raw

counts are shown

Press calibrate

softkey to calibrate

zero. Press menu

to cancel

calibrating.

Previous A/D raw

counts are shown

Press calibrate

softkey to calibrate

zero. Press menu

to cancel

calibrating.

Calibrating, please wait

Press enter to

remove offset from

zero and span

calibrations

POINT 1

POINT 2

POINT 3

POINT 4

POINT 5

WZERO

WVAL

WSPAN

REZERO

WLIN

WGT 1

CAL1

ID

DISPLAY

TEST

SETUP

AUDIT

CALIBR

SETPTS

Figure 3-6. Calibration Menu Layout

Table 3-5. Calibration Menu Parameters

CALIBR Menu

Parameter

Choices

Description

Level 2 submenus

WZERO

Press the calibrate softkey to display the AD raw counts.

Press calibrate softkey again to calibrate zero, or press MENU to cancel. “CALIBRATING” will
appear prior to automatically moving to WVAL.

WVAL

Press the calibrate softkey to display and edit the test weight value. Press the calibrate softkey
again to move to WSPAN.

WSPAN

Press the calibrate softkey to display the AD raw counts. Press the calibrate softkey to
calibrate the span or press MENU to cancel. “CALIBRATING” will appear prior to automatically
moving to WLIN.

WLIN

POINT 1 —
POINT 5

WGT 1 allows you to display and edit the test weight value. Press the calibrate softkey to edit
the value. CAL1 allows you to calibrate and display the raw AD value. Press TARE to perform
calibration. Press the calibrate softkey to move to the next calibration point.

Note

The linearity points are optional, but must not duplicate zero or span. They
must be between zero and span.

REZERO

Press the calibrate softkey to remove an offset value from the zero and span calibrations if
hooks or chains are being used during calibration.

Note

Use this parameter only after WZERO and WSPAN have been set. See
Section 4.1 on page 55 for more
information.