9.1

Error Messages

If an error code appears on the display, use the information in Table 9-1 as a troubleshooting guide. If you cannot
clear the error, call RLWS Service for assistance.

Table 9-1.

Counterpart

Error Messages

Error Display

Description

Solution

- - - - - -

Over range

• Check load cell wiring, including sense jumpers.
• Check configuration, including number of grads,

channel selection, display divisions.

• Check calibration, including W ZERO and

WSPAN values.

• Under range can be caused by OIML setting and

weight less than -20 display divisions.

• Check for scale binding or damage.
• Check for proper excitation voltage.
• Call RLWS Service for assistance.

- - - - - -

Under range

- - - - - -

- - - - - -

A/D out of range

CHKERR - BATTERY BACK LOST Battery Error

Press ENTER after replacing the battery to restore
defaults to battery backed storage.

ERROR

Internal program error

Check configuration. Run XE command (see
Section 9.2 on page 90) to determine error type. Call
RLWS Service if unable to clear error by cycling
power or if error reoccurs.

HWFERR

Hardware error

Reboot the unit. Call RLWS Service if unable to clear
error by cycling power or if error reoccurs.

INVALID UNITS

ID uses unassigned units

Modify the ID to use a configured unit.

NOTARE

Tare is prevented

Change regulatory mode settings or the TAREFN
parameter.

OVERFL

Overflow error

Weight value too large to be displayed.

OL

Overload error

Scale has been overloaded. Remove load from scale.

RANGE

GRADS > 100,000
WVAL > 100,000

Only shows up in Config mode.

EEPERR

EEPROM error

Call Rice Lake Weighing Systems (RLWS) for service

VERSION UPDATED

Core has been updated or memory
has been corrupted.

Press the ENTER key. If the message persists, call
RLWS for service.