Rice Lake Counterpart Configurable Counting Indicator - User Manual Rev. E - Version 2.5 User Manual

Page 60

Advertising
background image

54

Counterpart Counting Scale Indicator

Setpoint 1-6 sub-menu

GROSS
NET
PIECECNT
%REL

Value
Source
Trip
Bandval
Hyster
Preact
Preval
Preadj
Prestab
Pcount
Relnum
Batch
Pshaccm
Pshprint
Pshtare
Access
Digout
Sense

Display and edit the gross value

DELAY

Value
Source
Pshaccm
Pshprint
Pshtare
Access
Digout
Sense

COUNTER

Value
Access
Digout

Gross, Net, Peicecnt, %Rel, Delay and Counter sub-menus

VALUE

number

BANDVAL

number

HYSTER

number

Specifies a band around the setpoint value that must be exceeded before the setpoint, once
off, can trip on again.

PREVAL

number

PREADJ

0.500000
0-9999999

Preact adjustment factor. For setpoints with PREACT set to LEARN, specifies a decimal
representation of the percentage of error correction applied (0.05 = 50%, 1.0 - 100%) each
time a PREACT adjustment is made.

PRESTAB

0
0-65535

Preact stabilization time-out. For setpoints with PREACT set to LEARN, specifies the time, in
0.1-second intervals, to wait for standstill before adjusting the PREACT value. Setting this
parameter to a value greater than zero disables the learn process if standstill is not achieved in
the specified interval.

PCOUNT

1
0-65535

Preact learn interval. For setpoints with PREACT set to LEARN, specifies the number of
batches after which the preact value is recalculated. The default value, 1, recalculates the
preact value after every batch cycle.

RELNUM

1
2
3
4
5
6

For the percent relative setpoint, specifies the number of the relative setpoint. The target
weight for the %REL setpoint is determined as the percentage (specified on the VALUE
parameter of the %REL setpoint) of the target value of the relative setpoint.

Parameters

Choices

Description

Table 3-20. Setpoint Menu Parameters in Weigh Mode