elastomers). Use appropriate protective clothing and

safety goggles.

+

Interruption of the pump (e.g., due to power failure),

failure of connected components or of parts of the sup-

ply, or change in parameters must not be allowed to

lead to dangerous conditions. In case of a diaphragm

failure or in case of a leak in the manifold, pumped

substances might be released into the environment or

into the pump housing or motor.

Comply with all notes regarding proper use of the

pumps, as well as operation and maintenance guid-

ance.

+

The residual

leak rate of the equipment might render

possible an exchange of gas, albeit extremely slight,

between the environment and the vacuum system.

Adopt suitable measures to prevent contamination of

the pumped substances or the environment.

• Ensure that no parts of your clothing, hair or fingers

can be caught or drawn in at the inlet of the pump.

Never insert fingers or drop any other object into the

inlet or outlet.

• Pumping at high inlet pressure may lead to overpres-

sure at the gas ballast valve. Pumped gases or con-

densate might be expelled if the valve is open. If an in-

ert gas supply is connected to the gas ballast, ensure

that its inlet pipeline is not contaminated.

• You must take suitable precautions to prevent any

dangerous situation from arising if the controller starts

a vacuum pump (in combination with a VMS Module),

switches an in-line valve or a coolant valve, or opens a

venting valve.

Attention: If the controller is set to Autostart (Auto-
St.
: On

), the process will start immediately after a

power failure without pressing any further key. It is your

responsibility to ensure that automatic start-up of the

Advertising