system will not lead to any dangerous condition. Pro-

vide appropriate safety measures. Check prior to start-

ing the process whether the option ”Autostart” (device

menu) is enabled.

• Pay attention to the safety symbol ”hot surfaces” on

the equipment. Hot parts may cause burns if touched.

Adopt suitable measures to prevent any danger aris-

ing from hot surfaces or electric sparks. Ensure that

hot surfaces of the pump do not cause burns. Provide

a suitable contact guard if necessary.

• Ensure that the

coolant outlet pipe at the waste vapor

condenser is always free and that it cannot get blocked.

Prevent the backpressure of gases and the backflow of

condensates at the outlet.

Check the liquid level in the catchpots regularly and drain

condensate in time to prevent overfilling. Install a level

sensor (see ”Accessories”, pg. 74) for monitoring, if

necessary.

Provide appropriate protective measures to allow for the

possibility of failure and

malfunction. The protective mea-

sures must also allow for the requirements of the respec-

tive application.

The VNC 2 controller is powered by a short circuit proof

wide-range power supply with integrated overload protec-

tion and a user-servicable slow-blow fuse T8A (one spare

fuse included in fuse compartment of VNC 2).

In case of overload, the motor is shut down by a

self-hold

thermal cutout in the winding.

Note: Only manual reset is possible. Switch off the pump

and disconnect from the power source. Identify and elimi-

nate the cause of failure. Wait approx. five minutes before

restarting the pump.

NOTICE