Cleaning and inspecting the pump heads – VACUUBRAND PC 600 LAN NT User Manual

Page 83

Advertising
background image

page 83 of 102

The replacement of the diaphragm and the replacement of the valves can

be carried out separately.
+

To replace the valves, remove the head covers of one side of the pump

along with the assembled valve heads and fittings.

+

To maintain the diaphragms, the valve heads and the fittings need not

be disassembled. The head covers can be removed along with the as-

sembled valve heads and fittings.

+

Service only one side of the pump at a time to avoid the mixing of

parts.

Cleaning and inspecting the pump heads

Disassemble the exhaust vapor condenser

EK:

Loosen the union nut at the inlet of the ex-

haust vapor condenser.

Remove the 4 screws affixing the counter

holder of the exhaust vapor condenser with

a Torx driver T10. Remove the exhaust va-

por condenser. In doing so pull the PTFE

hose out of the inlet of the exhaust vapor

condenser.

Remove catchpots (9) at inlet and outlet

(see ”Use and operation”, pg. 31).

Loosen the union nut at the vacuum con-

nection of the controller.

Pull the PTFE tubing off the vacuum con-

nection.

Advertising
This manual is related to the following products: