Setting of the interface, Pin assignment rs-232 c – VACUUBRAND PC 600 LAN NT User Manual

Page 66

Advertising
background image

page 66 of 102

Setting of the interface

Set the interface parameters directly at the controller VNC 2. The factory

set values are underlined.

Edit the interface parameters in the device menu ”Select, RS 232”

using the s and t keys and confirm with ”Enter”.

Baud: 2400, 4800, 9600 or 19200

Parity: 7 data bits odd (Parity 7-O-1); 7 data bits even (Parity

7-E-1

); 8 data bits none (Parity 8-N-1)

Handshake (Handsh.): no Handshake (Handsh. None), XON/XOFF

Handshake (Handsh. Xon-Xoff), RTS/CTS Handshake (Handsh.
RTS-CTS

)

Remote On, Remote Off

Startbit = 1, Stopbit = 1

Timeout: Sending 1s, receiving 10s.

In remote mode (Remote On, ”P” is displayed) all keys at the controller

are inoperable.
To return to the manual operation of the controller, set the controller

to ”Remote off” via the interface, or switch off the controller. Then

switch the controller back on, and enter the device menu within 2s

by pressing the up or down key. Then select ”Select” / ”RS 232”

/ ”Remote” / ”Off”.

A maximum of ten commands per second is possible.

Read commands and commands ”REMOTE”, ”CVC”, and ”STORE”

can always be sent. The sending of other write commands is only pos-

sible, if ”Remote On” is selected.

The commands have to be written in capital letters.

Command and parameter have to be separated by a blank.

The string is terminated with <CR> or <CR><LF>.

The response of the controller is always terminated with <CR><LF>.

Numerical values and parameters can be written without leading zeros.

The response of the controller always includes leading zeros.

Pin assignment RS-232 C

2: RxD

5: Mass

9: +5V (Bluetooth)

3: TxD

7: RTS

4: DTR

8: CTS

Advertising
This manual is related to the following products: